Issue Date: December 21, 2005

PID Analyzers, LLC

(originally listed as HNU Systems, Inc.)

Model 102 Photoionizer

Out-of-Tank Product Detector (Vapor-Phase)

Detector

Output Type Quantitative
Sampling Frequency Intermittent
Operating Principle Photoionization Detector

Test Results

  Unleaded Gasoline Synthetic Gasoline JP-4 jet fuel
Accuracy* (%) - average reading 85 (423 ppm) 112.8 (1125.2 ppm) 8.1 (78.2 ppm)
Detection time* (sec) <120 11 12
Fall time* (sec) <120 38 11
Lower detection limit (ppm) Not Reported 2.1 5.8

*For tests conducted with 1000 ppm of test gas

Specificity Results

Activated: Activated: n-butane, n-hexane, iso-butane, 3-methylpentane, toluene

  • APID Analyzers, LLC
    25 Walpole Park South Drive
    Walpole, MA 02081
    Tel: (508) 660-5001
  • Evaluator: Ken Wilcox Associates
    Tel: (816) 443-2494
    Date of Evaluation: 10/29/04

Home |  Vendors: A–F |  Vendors: G–M |  Vendors: N–S |  Vendors: T–Z |  Method Index

Appearance on this list is not to be construed as an endorsement by any regulatory agency nor is it any guarantee of the performance of the method or equipment. Equipment should be installed and operated in accordance with all applicable laws and regulations. For full details, please refer to our expanded DISCLAIMER page.